	<entry>
		<id>https://newjune.uoregon.edu/mediawiki/index.php?title=Start-Up_Procedure&amp;diff=3597&amp;oldid=prev</id>
		<title>Aplstudent: Created page with &quot;&#039;&#039;&#039;Start-up Procedure:&#039;&#039;&#039;&lt;br&gt;  1. Make sure lab door is closed, and don appropriate eye protection.  2. Turn on power strip. (This should only turn on the UC-2000 controller,...&quot;</title>
		<link rel="alternate" type="text/html" href="https://newjune.uoregon.edu/mediawiki/index.php?title=Start-Up_Procedure&amp;diff=3597&amp;oldid=prev"/>
		<updated>2020-02-26T23:59:43Z</updated>

		<summary type="html">&lt;p&gt;Created page with &amp;quot;&amp;#039;&amp;#039;&amp;#039;Start-up Procedure:&amp;#039;&amp;#039;&amp;#039;&amp;lt;br&amp;gt;  1. Make sure lab door is closed, and don appropriate eye protection.  2. Turn on power strip. (This should only turn on the UC-2000 controller,...&amp;quot;&lt;/p&gt;
&lt;p&gt;&lt;b&gt;New page&lt;/b&gt;&lt;/p&gt;&lt;div&gt;&amp;#039;&amp;#039;&amp;#039;Start-up Procedure:&amp;#039;&amp;#039;&amp;#039;&amp;lt;br&amp;gt;&lt;br /&gt;
&lt;br /&gt;
1. Make sure lab door is closed, and don appropriate eye protection.&lt;br /&gt;
&lt;br /&gt;
2. Turn on power strip. (This should only turn on the UC-2000 controller, everything else should still be off).&lt;br /&gt;
&lt;br /&gt;
3. On UC-2000 Controller:&lt;br /&gt;
* Hold the &amp;#039;SELECT&amp;#039; and &amp;#039;ENTER&amp;#039; buttons down to enter &amp;#039;Setup Mode&amp;#039;.&lt;br /&gt;
* Set Frequency to 5 kHz, Gate Pull to &amp;#039;Up&amp;#039;, Maximum PWM to 95%, Lase On Pwr-Up to &amp;#039;N&amp;#039;, Checksum to (On). Use the &amp;#039;SELECT&amp;#039; &amp;lt;br&amp;gt;&lt;br /&gt;
: button to switch between options and the &amp;#039;ENTER&amp;#039; button to change the setting selected. Once finished, select &amp;#039;SAVE AND EXIT&amp;#039; &amp;lt;br&amp;gt;&lt;br /&gt;
: and hit the &amp;#039;ENTER&amp;#039; button on the controller.&lt;br /&gt;
* Set mode to manual by hitting the &amp;#039;SELECT&amp;#039; button on the main screen until &amp;#039;MANUAL&amp;#039; option comes up, and hit &amp;#039;ENTER&amp;#039;. &amp;lt;br&amp;gt;&lt;br /&gt;
: (Make sure not to select &amp;#039;MANUAL CLOSED&amp;#039;, as this may not operate in output intended if not in the recommended range.)&lt;br /&gt;
* Turn adjustment knob until PWM is set to 0.0%.&lt;br /&gt;
&lt;br /&gt;
4. Turn on power supply unit for cooling fans. Turn coarse voltage to ~12V, and current to ~0.5A. (The maximum power rating of each&lt;br /&gt;
:is 24V @ 0.46A each, and can be increased as needed if laser begins to heat during operation.)&lt;br /&gt;
&lt;br /&gt;
5. Turn on power supply for Laser. Set voltage to 30V and turn knob for current all the way to the right. The current will not appear&lt;br /&gt;
:to rise at this point, but when the laser is turned on it will draw as much current as necessary for the set PWM %.&lt;br /&gt;
&lt;br /&gt;
6. Remove aperture dust cover.&lt;br /&gt;
&lt;br /&gt;
7. Set PWM to 1.0% and toggle &amp;#039;LASE&amp;#039; to &amp;#039;ON&amp;#039;. You can confirm the laser is in operation by the red LED being lit on the rear of the&lt;br /&gt;
:laser. Using an IR Detector card, ensure beam is aligned with an appropriate beam block or power meter, ensuring the beam does&lt;br /&gt;
:not go past the table.&lt;br /&gt;
&lt;br /&gt;
8. Once beam is safely aligned, this is now ready to use.&lt;/div&gt;</summary>